Specifications
IMPEDANCE MEASUREMENT
Type: Class 2 acc. to ANSI S3.39
TYMPANOMETRY
Test frequency: 226, 678, 800, 1000 Hz ± 2%
Test level: 85 dBSPL in 2 cm3 at 226 Hz; 85 dBSPL at 678, 800, 1000 Hz
Pressure range: Automatic: +200 to -400 daPa
Manual: +300 to -600 daPa
Volume range: 0.1 to 6.0 ml
Accuracy: ± 10 % or ± 10 daPa
Test time: Less than 3 seconds
Compliance range: 0 to 3.0 ml